OZONE MONITOR BMT 932 Manual, Rev. 02/2021
45
Cmd Meaning PC->932 932->PC
47 Set Alarm Beep
1: On
0: Off
*47#byte *47#
48 Start Scrubbertest *48# *48#
49 Send Error Log (max. 16 Entries)
Format: YY,MM,DD,HH,MM,SS,Error
Error has the same format as described in table on page 30
*49# *49#byte1,
byte2,byte3,
byte4,byte5,
byte6,word#...
50 Send Event Log (max. 103 Entries)
Format: YY,MM, DD,HH,MM,SS, Word, Float
The 16 bits in word have the following meaning and are sent with the
following float parameter:
*50# *50#byte1,
byte2,byte3,
byte4,byte5,
byte6,word,float#
byte1,...
7-Bits Meaning float Parameter
0 Low Alarm status Channel 1 Concentration
1 Low Alarm status Channel 2 Concentration
2 Low Alarm status Channel 3 Concentration
3 Low Alarm status Channel 4 Concentration
4 Low Alarm status Channel 5 Concentration
5 Low Alarm status Channel 6 Concentration
6 Switched on pressure [bar]
7 Switched off Temperature when
switched on [K]
8 High Alarm status Channel 1 Concentration
9 High Alarm status Channel 2 Concentration
10 High Alarm status Channel 3 Concentration
11 High Alarm status Channel 4 Concentration
12 High Alarm status Channel 5 Concentration
13 High Alarm status Channel 6 Concentration
14 High Alarm, any Channel Concentration
15 Low Alarm, any Channel Concentration
54 Reset all parameters to standard factory setting *54# *54#
66 Active Channels?
Shows which channels are active, returns 0..5 for channels 1..6
*66# *66#byte
67 Set Active Channels
Bits 0..5 stand for channels 1..6. This may never be zero
*67#byte *67#
68 Low Alarm EMO-Mask 1?
Reads out assignment of Low Alarms to EMO 1, bits 0..5 set show
assignment of Low Alarms on channel 1..6
*68# *68#byte
69 High Alarm EMO-Mask 1?
Reads out assignment of High Alarms to EMO 1, bits 0..5 set show
assignment of high Alarms on channel 1..6
*69# *69#byte
73 Low Alarm EMO-Mask 2?
Reads out assignment of Low Alarms to EMO 2, bits 0..5 set show
assignment of Low Alarms on channel 1..6
*73# *73#byte
74 High Alarm EMO-Mask 2?
Reads out assignment of High Alarms to EMO 2, bits 0..5 set show
assignment of high Alarms on channel 1..6
*74# *74#byte
75 Low Flow Alarms?
Shows Low Flow Alarms as bits 0..5 for channel 1..6
*75# *75#byte