10 Storage
38 ARC 100 Operating Manual 900-100_IFU-V1.0_11587-S0-20141209-EN
9.3. Technical service
Contact the following service centre for maintenance and repair work:
BOWA-electronic GmbH & Co. KG
Heinrich-Hertz-Strasse 4-10
72810 Gomaringen, Germany
Phone +49 (0) 7072-6002-0
Fax +49 (0) 7072-6002-33
E-mail: service@bowa.de
or visit our website:
www.bowa-medical.com
10. Storage
Pos: 71 /679-BOW A/HF-Gerät e/ARC 300/350/1 0 Lagerung/L agerung @ 4\ mod_12610407 15569_258.doc @ 35279 @
If you store the HF device longer than one year, pay particular
attention to the indicators of the automatic function tests; see
Section Functional test, page 28.
Clean the HF device thoroughly before you put it into storage.
Store the HF device in a clean, dry place in accordance with the
storage conditions.
Storage conditions:
Temperature: -10 °C to +60 °C
Relative humidity 10 % to 85 %,
non-condensing
Atmospheric pressure: 500 hPa to 1060 hPa