What to do if the Bowa Medical Equipment smoke evacuation system is ON but suction is minimal or none?
- RRyan RoseAug 2, 2025
If the smoke evacuation system has minimal or no suction, there could be several reasons: * The filter may not be seated correctly. Re-install the filter, pressing firmly into place and fully engage the clip. * The filter might be clogged. Replace the filter with a genuine BOWA filter. * The vacuum hose or tube may be clogged. Replace the vacuum hose or tube with genuine BOWA products. * The motor/blower may be obstructed. In this case, call BOWA Technical Services.