MN009-940-S0 EN BOWA SHE SHA smoke evacuation unit 17
2.5 Electromagnetic Compatibility Information per IEC60601-1-2
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Surgical Smoke Plume Evacuator model SHE SHA is intended for use in the electromagnetic
environment specified below. The customer or user of the SHE SHA should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The SHE SHA uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A
The model SHE SHA is suitable for use in all
establishments, other than domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Not applicable.
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Class A
Not applicable.