MN009-940-S0 EN BOWA SHE SHA smoke evacuation unit 23
3.5 Troubleshooting the System see below.
Problem
Potential Cause
Corrective Action
1. Smoke Evacuation
System is ON but
suction is minimal
or none.
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum hose or tube
is clogged.
4. Motor/blower is
obstructed.
1. Re-install filter, press firmly into place and fully
engage clip.
2. Replace filter with a genuine BOWA filter.
3. Replace vacuum hose or tube with genuine
BOWA products.
4. Call BOWA Technical Services
2. Smoke Evacuation
System does not
function even though
suction ON button is
depressed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic system
failure.
4. Filter life has expired
or invalid filter
inserted.
1. Check power outlet and connection to rear or
side panel of the machine.
2. / 3. Call BOWA Technical Services
4. Replace filter with a genuine BOWA SHE SHA
filter.
4. Technical Service
Contact the following service center for maintenance and repair work:
BOWA-electronic GmbH & Co. KG
Heinrich-Hertz Straße 4-10
72810 Gomaringen, Germany
Phone: +49 (0) 7072-6002-0
Fax: +49 (0) 7072-6002-33
E-Mail: service@bowa.de