MN009-940-S0 EN BOWA SHE SHA smoke evacuation unit 5
1.2 Inspection
The SHE SHA Smoke Evacuator has been thoroughly tested and inspected before shipment from
the factory. Please check the unit before using it to insure that no damage has occurred in transit. If
damage is evident, please contact the Technical Service, see chapter 4 technical service.
In addition, please compare the accessories you receive with the standard accessories list below. If
an item is missing, please notify BOWA Customer Service.
Standard Accessories:
Operating Manual
Power Cord
Pneumatic Foot switch
Remote Switch Activator
Please contact BOWA Customer Service to purchase the following accessories:
Replacement Filters
Hoses, Tubing, Adapters, Handles & Other Accessories
1.3 Operational Information
The operational information contained in this section is intended for the customer review of
regulatory issues. The information pertains to the use of the products both domestically and
internationally:
1. The BOWA SHE SHA Smoke Evacuation System(s) complies with IEC60601.1 electrical
specifications in the following systems:
220/240 VAC 50/60 Hz
2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I
3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part
4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1
5. Method of sterilization or disinfection recommended by BOWA (UL 60601-1, Clause 5.4):
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.
Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable
7. Mode of operation (UL 60601-1, Clause 5.6): Continuous
8. Upon request, BOWA will provide the following:
Service and Repair Instructions, including Circuit Diagrams and Parts List
9. The fuses on the circuit board are to be serviced by an authorized BOWA technician as follows:
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-Blo), (F1,F2)