5 Preparation
34 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
5. Preparation
Pos: 31 /679-BOW A/HF-Geräte/ ARC 300/350/5 Vorbereitung/HF- Gerät aufstelle n @ 6\mod_ 12988737536 90_6.doc @ 490 04 @ 2
5.1. Setting up the HF device
HF devices may be used only in rooms used for medical purposes that
meet the requirements of DIN VDE 0100-710.
If the HF device was previously stored or transported at temperatures
below +10 °C, it will take approximately three hours to adjust to room
temperature.
1. Observe and comply with the operating conditions, see
chapter Operating conditions, page 33
2. Place the HF device on one of the following platforms:
a table,
an equipment trolley,
a console suspended from a ceiling support or wall-
mounted brackets.
if necessary, position the HF generator on the additional
device ARC PLUS
Except for the additional device ARC PLUS, do not place
any other devices on the HF device