7 Detecting and correcting faults
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 53
No power set for the activated output
Set the power for the activated output.
Connection socket activated via finger
and foot switches
At the end of activation a different
activation signal is pending for the
same socket
Activate the output via the finger or
foot switch one after the other.
Tissue contact is already pending
upon activation of the AUTOSTART
mode
Remove the instrument from the
tissue and re-activate AUTOSTART.
Papillotome or snare is not applied
Apply the papillotome or the snare.
Instrument for GastroCut application
defective or short circuit
Replace the Gastro instrument.
Select a new point of application.
Excessive low-resistance activation
(e.g. in saline)
Ensure that the instrument has
sufficient contact with the tissue for
coagulation or cutting before
activating the device.
EASY neutral electrode error
Check the neutral electrode and the
neutral electrode cable, see
section Neutral electrode
monitoring, page 30.
Restart the device. If the error recurs,
contact the service center.
Restart the device. If the error recurs,
contact the service center.