11 Technical specifications
70 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
Setting
Program
Gastro LOOP
6
Values measured on ohmic resistances
Output power P [W] versus rated load
resistance R [Ω] with "Gastro LOOP",
"Blend 09" settting, COAG phase
Values measured on ohmic resistances
Output power P [W] versus load
resistance R [Ω] with "Gastro LOOP",
"Blend 09" setting, CUT phase
Blend
Coag share
in W
Cut share
in W
0
36
370
1
36
370
2
36
370
3
36
370
4
36
370
5
36
370
6
36
370
7
36
370
8
36
370
9
36
370
Output power P [W] versus Blend
setting for "Gastro LOOP". Rated load
resistance: Coag 75 Ω, Cut 500 Ω
Blend
HF voltage [Vp]
0
800
1
800
2
800
3
800
4
800
5
800
6
800
7
800
8
800
9
800
HF output voltage U [Vp] versus
"Gastro LOOP" "Blend 09" (no load)
The short-term peak power exceeds 400 W, but this is permissible
because the mean value over 1 second is below 400 W.