11 Technical specifications
70 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
Values measured on ohmic resistances
Output power P [W] versus rated load
resistance R [Ω] with "Gastro LOOP",
"Blend 0–9" settting, COAG phase
Values measured on ohmic resistances
Output power P [W] versus load
resistance R [Ω] with "Gastro LOOP",
"Blend 0–9" setting, CUT phase
Output power P [W] versus Blend
setting for "Gastro LOOP". Rated load
resistance: Coag 75 Ω, Cut 500 Ω
HF output voltage U [Vp] versus
"Gastro LOOP" "Blend 0–9" (no load)
The short-term peak power exceeds 400 W, but this is permissible
because the mean value over 1 second is below 400 W.