250 - 450 Sonifier Chapter 6: Operation
User’s Manual Operating Considerations
100-413-016 Rev. C 6 - 7
You can effectively treat all but the most difficult of materials by the following
two steps:
1. Homogenize the tissues or solids by placing them in a high-speed
blender with the solution.
2. Insert the horn in the liquid sample for complete disruption.
To disrupt solid pieces, especially those that are extremely resistant to breakage,
place the horn directly over the tissue or right against it.
6.2.6 Using Glass Powders with Solution
To disrupt difficult cells and tissues, adding glass powders (5 microns to 0.5 mm)
will materially decrease treatment tissues, especially when used in conjunction
with the standard, high intensity micro tip. A ratio of 1 part glass powder to 2 parts
liquid is recommended.