Air bubble/accumulated air - Respond “Yes” or “No”
to prompt for priming to clear the air. If “Yes” is selected
to prime IV line, respond to prompt to disconnect IV line
from patient.
Danger of Free Flow - Anti-free ow clip is not prop-
erly inserted. Conrm danger free ow messages, clamp
the line, respond to open door prompt, remove the IV
line and reinsert IV line per IV line loading instructions.
Other TROUBLESHOOTING
Door fails to close - Press open door button again
while door is open and re-close door.
Persistent air in line alarm with no visible air -
Close roller clamp, open door, remove IV line and inspect
for bubble trapped in IV line at air in line sensor.