WIRELESS DRUG LIBRARY UPLOAD
1. Observe alternating with on the top line
of the pump display when a new Drug Library is
available.
2. Press
to stop the infusion. Follow instructions above
under “End Infusion and Power O” to power o.
3. Wait 10 seconds, a progress bar will appear on
pump, do not power pump back on until Drug Library
Upload is complete as indicated by progress bar.
4. When pump is re-started, respond to two prompts
that all values are cleared and new drug library has
been loaded.
RED ALARMS AND TROUBLESHOOTING
Red alarms stop the infusion. The pump emits an audible
signal and red LED ashes. The display states “Alarm”
and the reason for the alarm. The alarm signals are
conrmed and turned o by pressing . Correcting the
alarm state depends on the cause of the alarm.
Red alarms:
“VTBI infused” “Time expired”
“Battery empty” “Downstream Occlusion”
“KVO nished” “Standby time expired”
“Upstream Occlusion” “Air bubble/Accumulated air”
“Danger of Free Flow”
Troubleshooting RED ALARMS
Downstream/Upstream pressure - Check clamps and
patient line.