OPEN DOOR AND LOAD IV LINE
1. Press
to power on.
2. Press door open key
. Respond to prompt and
press to open door.
CAUTION: Keep the roller clamp closed while inserting
IV line.
3. (1) Load IV line from right to left. Insert the
two hole clip.
(2) Attach white pronged clip.
NOTE: Ensure silicone tubing segment is not twisted.
(3) Insert the green anti-free ow clip in the green
slot (with hook facing down and away from pump)
until the opening lever snaps out and the ashing
yellow light goes out indicating pump has occluded
IV line.
4. Press IV line into the air sensor guide and notches
on right and left side of the pump.
5. Close the pump door by rmly placing pressure until
you feel the pump pull the door shut.
NOTE: Do not open the roller clamp until programming
prompt appears.
CAUTION: Pump door will not close and pump alarms
if the anti-free ow clip is not properly inserted. Do not
force door closed, check IV line and anti-free ow clip
(green) for proper loading.