4 5
10. Use and to move to the digit to be
programmed.
NOTE: The
may be used to clear existing values.
11. Use and to program new values.
12. Press to conrm value.
13. If value is outside soft limits, respond to prompt with
for “Yes” to override soft limit and for “No”.
Note change in soft limit symbol when overridden.
or .
NOTE: If value is outside hard limit, press
to conrm
message and edit value.
14. Scroll down to enter and/or review VTBI and Time.
VTBI is required to start the infusion.
15. Press
to begin infusion.