SECONDARY INFUSION
NOTE: Prime SECondary IV line manually, close roller
clamp and connect to PRIMary IV line at needleless port
ABOVE the pump.
1. Stop PRIMary infusion by pressing
.
2. Scroll to and select “SECondary” by using
.
3. Four sub-menus will be displayed:
• New SECondary
• Use a Previous SEC infusion
• Back to PRIM
Determines switch back to PRIMary. “manual”
stops the pump and alarms when the SECondary is
complete. “Auto-change” converts to the PRIMary
infusion when the SECondary is complete.
NOTE: The method of switching back to PRIMary
remains for all SECondary infusions until changed.
• Basic infusion
4. Select “New SECondary” with
.
5. Select drug using
and conrm with .
NOTE: SECondary medications will most frequently be
set up as Dose Over Time infusions.
6. Enter Total Dose using
and conrm with .
7. Enter Total Time using
and conrm with .
NOTE: The pump calculates rate and VTBI.
8. Press
, note message to check bag height
and unclamp SECondary. Press again to conrm
message and begin infusion.