6 7
NOTE: The SECondary bag must be hung at least 8
inches above the PRIMary, measured from the bottom
of the SECondary bag to the top of uid level in the
PRIMary.
NOTE: If SECondary rate exceeds 300 ml/hr clamp o
PRIMary to prevent concurrent ow.
9. To resume PRIMary prior to completion of SECondary,
press
. Scroll to and select “resume PRIMary”
using arrow keys . Conrm message “Infusion
mode changed to PRIMary” with .
10. Press . Note message to clamp or remove
SECondary. Press again to infuse PRIMary.
NOTE: To resume SECondary, follow same steps as above.
NOTE: To clear a SECondary infusion prior to
completion, press
, then Press and to
“delete SECondary”.
NOTE: Pump stores 5 of the previously administered
SECondary medications. They are cleared when the Care
Unit changes or the pump is powered o.
11. To select a previously administered SECondary,
stop PRIMary by pressing
.
12. Scroll to and select “SECondary” menu using
.
13. Scroll to and select “Use a previous SEC infusion”
using
.
14. Scroll to and select drug using
.
15. When all values are checked, press
and then
press
.