8 9
LOADING DOSE
1. Press
to answer yes to loading dose prompt.
2. Enter loading dose amount using
and conrm
with .
3. Enter loading dose time using arrow keys
and
conrm with .
4. Press
to access doserate/rate editor to program
continuous infusion.
5. Enter doserate or rate for continuous infusion using
and conrm with .
NOTE: “LDdos” and “Start” alternate in upper right
corner of screen.
6. Press to begin infusion.
7. The loading dose may be stopped at any time by
pressing . Press to conrm prompt with
amount delivered of total programmed.
8. Press to resume infusion, respond to prompt
to continue loading dose with
for “Yes” and
for “No”.
NOTE: Once “No” is selected, continuous infusion will
begin and loading dose is no longer accessible.
BOLUS DOSE
1. Press .
2. If prompted, select “Program a bolus dose and time”
using
and select with .
3. Enter bolus dose using
and conrm with .
4. Enter bolus time using
and conrm with .
5. Review bolus parameters and press
to begin
bolus.