Cleaning and disinfecting the Stimuplex
®
HNS 11
Use only soft, damp cloths to clean and disinfect the Stimuplex
®
HNS 11 and the electrode cable. Water, soap or white spirits are
especially suitable for this. Ensure that no moisture penetrates the
Stimuplex
®
HNS 11.
Use only wet disinfection, no spray disinfection! Avoid
condensation!
White spirit or commercially available methanol-free disinfectant
in an ethyl alcohol base can be used for disinfection.
Caution!
The following substances must not be used:
trichloroethylene, acetone, methyl ethyl ketone, benzene, metha-
nol, cellulose thinner, 1-propanol and 2-propanol.
Maintenance and safety checks
Check the serviceable condition of the Stimuplex
®
HNS 11 and ac-
cessories on each occasion before use. A defective instrument must
not be used.
Electromedical equipment may only be repaired by the manufac-
turer or by an organization expressly authorized by the manufac-
turer. An order for repairs must be accompanied by a detailed
description of the fault (acc. separate Technical Service Manual).