Product specifications
Model: IRT 3030
Displayed temperature
range: 93.2 °F – 108 °F (34 °C – 42.2 °C)
Operating ambient
temperature range: 50 °F – 104 °F (10 °C – 40 °C)
Storage temperature: -25°C - 55°C (-13°F - 131°F)
Display resolution: 0.1°F or 0.1°C
Accuracy for displayed
temperature range:
±0.2°C in range 35-42°C (95-107.6°F)
±0.3°C outside that temperature range
Clinical bias:
-0.198 °F (-0.11°C)
Limits of agreement:
±
0.45 °F
(
0.25 °C)
Clinical repeatability:
±
0.18 °F (0.10 °C)
Measuring Site: Ear
Reference Site: Oral (estimated oral temperature
is displayed)
Operating and storage
relative humidity: ≤ 95% non-condensing
Battery life: 2 years /1000 measurements
Service life: 5 years
Warranty: 3 years
If device is not used or stored within specied temperature and
humidity ranges the technical accuracy cannot be ensured.
This infrared thermometer meets requirements established in
ASTM Standard E 1965-98 (for the thermometer system
[thermometer with lens lter]). Full responsibility for the
conformance of the product to the standard is assumed by
Kaz USA, Inc.
ASTM laboratory accuracy requirements for the thermometer
only (not including a lens lter) in the display range of 96.8 °F
to 102.2 °F (36 °C to 39 °C) for infrared thermometers is ± 0.4 °F
(± 0.2 °C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86
and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).