6
1. To turn the thermometer on, push the “start” button (4).
During an internal self-check, the display
shows all segments.
The reading showed on the screen is the
last temperature taken.
Flashing symbol of lens lter will appear if
no lens lter is attached. Temperature can
be taken only if lens lter is mounted on
the probe.
The thermometer is ready to use after two
beeps are sounded and the non-ashing
ear icon appears.
2. Fit the probe (3) snugly into the ear canal,
then push the “start” button (4).
One beep signals normal temperature.
Audio temperature indicator will trigger 2
consecutive beeps when the reading is
equal or higher to 99.5 °F (37.5 °C) and 4
beeps if the reading is equal or higher to
101.5 °F (38.6 °C). The result is shown on
the display (7).
3. Gently pull o the disposable lens lter (2).
Do not use force in removing the
disposable lens lter.
4. Put on a new, clean, disposable lens lter by
snapping into position.
5. Place the protective cap on (1).
The Braun ThermoScan® 3 High Speed
Compact Ear Thermometer turns o
automatically after 60 seconds of inactivity.
The thermometer can be turned o by pressing
and holding the start button for more than 5 seconds.