Changing the temperature scale
Make sure the thermometer is turned o. Press and
hold down the start button. Keep holding down
the “start” button and after about 3 seconds the
display will start to cycle between temperature
scales (°F or °C).
Release the start button when the desired scale is
shown. This will reset the thermometer to the new
temperature scale.
Replacing the battery
The thermometer is supplied with one 3V Lithium cell battery (type
CR2032). Insert new battery when the low battery symbol appears
on the display.
The battery door design is
intended to prevent a small child
from easily opening the battery
door to address a possible choke
hazard. Open the battery
compartment by inserting a
narrow pointed object such as a ball point pen or paperclip into the
small hole above the battery door (8). While pressing into the hole
slowly slide back the battery door with the other hand.
To close battery door align the tabs along the side of the battery
into the battery compartment and slowly slide back into position
until the door clicks into its locked position.
Do not mix alkaline, standard, or rechargeable batteries.
Calibration
This product has been calibrated at the time of manufacture. If
used according to the use instructions, periodic calibration is not
required. If at any time you question the accuracy of temperature
measurements, please contact Consumer Relations.