13
English
Displayed temperature range:
93.2 °F – 108 °F (34 °C – 42.2 °C)
Operating ambient
temperature range: 50 °F – 104 °F (10 °C – 40 °C)
Storage temperature: -25°C - 55°C (-13°F - 131°F)
Display resolution: 0.1°F or 0.1°C
Accuracy for displayed
temperature range: ±0.4 °F (±0.2 °C) in range 35.5-42°C
(95.9-107.6°F)
±0.5 °F (±0.3 °C) outside that temperature
range
Clinical accuracy characteristics and procedures are available
upon request.
Measuring Site: Ear
Reference Site: Oral (estimated oral temperature
is displayed)
Operating and storage
Relative humidity: ≤ 95% non-condensing
Battery life: 3 years /1000 measurements
Service life: 5 years
Warranty: 3 years
If device is not used or stored within specied temperature and
humidity ranges the technical accuracy cannot be ensured.
This infrared thermometer meets requirements established in
ASTM Standard E 1965-98 (for the thermometer system
[thermometer with lens lter]). Full responsibility for the
conformance of the product to the standard is assumed by Kaz
USA, Inc.
ASTM laboratory accuracy requirements in the display range of 37
to 39°C (98 to 102°F) for IR thermometers is ±0.2°C (±0.4°F),
whereas for mercury-in-glass and electronic thermometers, the
requirement per ASTM Standards E667-86 and E1112-86 is ±0.1°C
(±0.2°F).