Braun ThermoScan PRO 3000 - Troubleshooting

Braun ThermoScan PRO 3000
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
message
Troubleshooting
Situation
Solution
No
probe
cover
is
attached
and
activation
button
was
pressed
while
probe
was
in
the
ear.
Make
sure
probe
window
is
clean;
refer
to
section
«Care
and
cleaning»
if
necessary.
Attach
new,
clean
probe
cover
to
stop
error
beeps.
Ambient
temperature
is
not
within
the
allowed
operating
range
of
50°F
to
104°F
(10°C-40°C)
or
changing
too
rapidly.
Allow
the
thermometer
to
remain
in
a
room
for
30
minutes
where
the
temperature
is
between
50°F
to
104°F
(10°C-40°C).
Temperature
taken
is
not
within
displayed
temperature
range
(68°F
-108°F
/20°C-
42.2°С).
HI
=
too
high
LO
=
too
low
Make
sure
new,
clean
probe
cover
is
attached
and
thermometer
is
properly
inserted.
To
clear
the
error
message,
attach
a
new,
clean
probe
cover
or
press
the
©/mem
buiton.
Then,
take
a
new
temperature.
Battery
is
low,
but
thermometer
will
still
operate
correctly.
(No
light
function)
Insert
new
batteries
Battery
is
too
low
to
take
temperature
readings.
Insert
new
batteries
System
error
e
If
error
persists
e
If
error
still
persists
Wait
30
seconds
until
thermometer
powers
down,
then
turn
on
again.
...
reset
the
thermometer
by
removing
the
batteries
and
putting
them
back
in.
...
Call
Braun
ThermoScan
Customer
Service
at
1-800-327-7226.
11