14
Product
specifications
Displayed
temperature
range
68
°F
-
108°F
(20°C
-
42.2°C)
Operating
ambient
temperature
range
50
°F
-
104
°F
(10°C
-
40°C)
Display
resolution
0.1°F
or
°C
Accuracy
for
patient
temperature
range
Maximum
Laboratory
Error
96.8
°F
-
102.2
°F
(36°C
-
39°C):
+0.4°F
+0.2°C
outside
this
range:
+0.5°F
+0.3°C
Long
term
storage
ranges
Temperature:
~4°F
to
122°F
(-20°C
to
50°C)
Humidity:
95%
non-condensing
Shock:
Withstands
one
meter
drop
Measurement
time
1
second
Minimum
time
between
readings
2
seconds
Automatic
«power
down»
30
seconds
Battery
life
5,000
temperatures
Battery
type
2 x
CR
2032/DL
2032
Thermometer
dimensions
6.4"
x
1.7"
x
1.2"
(16.3
cm
x
4.4
cm
x
3.1
cm)
Thermometer
weight
3.5
ounces
(100
g)
{without
batteries)
This
infrared
thermometer
meets
requirements
established
in
ASTM
Standard
E
1965-98.
Full
responsibility
for
the
conform-
ance
of
the
product
to
the
standard
is
assumed
by
Braun
GmbH,
61476
Kronberg,
Germany.
ASTM
laboratory
accuracy
requirements
in
the
display
range
of
96.8
°F
to
102.2
°F
(36
°C
to
39°C)
for
infrared
thermometers
is
+0.4
°F
(+0.2°C),
whereas
for
mercury-in-glass
and
electronic
thermometers,
the
requirement
per
ASTM
Standards
E667-86
and
E1112-86
is
+0.2
°F
(+0.1°C).