Brookeld Engineering Labs., Inc. Page 10 Manual No. M04-242-E1213
CAUTION: When using the1/2pintcan,donotlowerthespindledirectlyintothe
container.The narrowdiameterof thecanrequires thespindleto be
introducedatanangle.Tiltthe1/2pintcanwhileloweringtheviscometer.
8) Thespindlewillbegintorotateoncethehandleiswithin1/2inchofthelowestposition.
9) Wait5secondsforthedisplayreadingtostabilize.Adisplayof"----"indicatesanover-range
condition.Ifaprinterisbeingused,thesedasheswillprintinplaceofdata(Figure7).
10) Movethe Hold Reading Switch to the down position
to"Hold"thedisplay.Ifaprinterisconnected,astring
ofdatawillprintassoonastheHoldreadingswitchis
movedtothedownposition(Figure7).Thedisplayunit
selectorswitchmaybeusedtochangethedisplay.
11) Raisethehandletothetopposition.Thisstopsthespindle
fromrotating.
Note:Whenusing1/2pintcans,youmustliftthecanandthentiltslightlytoremovethe
spindle.
12) Loosenthethumbscrewandremovespindleforcleaning.
Note:Donotremovethethumbscrew.Ifdropped,itmayfallintothesample.
II.3 Troubleshooting
IftheoperatinghandleisintheDOWNpositionandthedisplaylocksupand/orthespindledoes
notrotatethefollowingshouldbechecked:
• BesuretheHOLDswitchisintheUPposition.IftheholdswitchwasintheDOWNposition
priortoengagingtheoperatinghandle,dashes(——)orthepreviousreadingmayappearin
thedisplay.
• TheKU-2requiresashutdown(poweroff)periodofatleast30secondsbeforeitcanbeoperated
properly.Iftheunithadpreviouslybeenturnedoffandimmediatelyturnedbackon(without
a30secondshutdownperiod)boththedisplayandrotationmaytemporarilylockup.
Ifyouexperiencetheseproblems,simplyraisetheoperatinghandle,turntheunitoffandwaita
minimumof30secondsbeforeturningthepowerbackon.
Figure 7
82g 55.9KU 272cP
----g ---.-KU ---cP