Icon System Manual T02985_02_03  11
 
MRI System Overview 
 
 
The  ICON  magnet  has  integrated  magnetic  shielding.  Laterally,  the  0.5 mT  line  is 
protruding the magnet surface < 40 cm (14 inches). On top of the magnet the 0.5 mT line 
is completely inside the magnet housing. 
 
 
Risk of temporary impairment for persons with cardiac pacemakers. 
Cardiac pacemakers could temporarily switch into test mode in the vicinity of the 
magnet. 
  Persons  with cardiac pacemakers must not approach  the magnet closer than 
35 cm / 14 inches. 
2.2  Gradient and shim coils 
The ICON MRI magnet has a built-in set of three magnetic field gradient coils (x, y, z).  
The gradient coils are cooled by air-flow through the magnet. The air-cooling fans may be 
seen in Figure 2.2 (3). Never cover these to allow enough air flow through the gradient 
coils. 
For  temperature  monitoring  of  the  gradient  coil  two  temperature  sensors  are  built  in.  The 
temperature  supervision  uses  these  sensors  to  automatically  shut  down  the  gradient 
amplifier  whenever  a  threshold  temperature  of  approximately  45°  Celsius  inside  the 
gradient coils is exceeded. 
To protect the gradient from damage by duty cycle violations, the actual power applied to 
the gradient coil is monitored and the gradients are automatically turned off as soon as the 
power is over the allowed limit. 
The  gradient  coils  are  also  used  as  “shim”  coils,  i.e.  for  transiently  optimizing  the 
magnetic field homogeneity when inserting a sample or an animal. The ICON  system 
does not feature second or higher order shim coils. 
The gradient coils are virtually eddy-current free meaning that gradient shapes are near 
perfect without using electronic or digital pre-emphasis.