Questions and Answers

  • J
    Johnny MossSep 12, 2025
    What to do if shims do not affect the NMR signal on Bruker 700'54 Ascend Laboratory Equipment?
    • C
      Chad HuntSep 12, 2025
      If shims do not affect the NMR signal in your Bruker Laboratory Equipment, it could be because the shim heater current is not correct, and the shim switches are not opened. Set the shim heater current to the specified value.
  • N
    nathan48Aug 25, 2025
    What to do if Tilt Watch or Shock Watch is activated on Bruker Laboratory Equipment?
    • J
      Joe RileyAug 25, 2025
      If the Tilt Watch or Shock Watch on your Bruker Laboratory Equipment is activated due to careless transportation, accept the delivery with reservation, remark the extent of the damage in the transportation documents, and start a complaint process.
  • C
    Corey TorresAug 25, 2025
    What to do if there is visible damage on Bruker 700'54 Ascend?
    • S
      Samantha GarciaAug 26, 2025
      If you notice visible damage to your Bruker Laboratory Equipment due to careless transportation, accept the delivery with reservation. Make sure to note the extent of the damage in the transportation documents and initiate a complaint process.
  • P
    Patricia PerezSep 23, 2025
    What to do if there is high helium flow after breaking vacuum in Bruker 700'54 Ascend Laboratory Equipment?
    • A
      abbottkathleenSep 23, 2025
      If there is high helium flow after breaking vacuum in your Bruker Laboratory Equipment, it means that there are remaining cryogenic agents in the vessels. Remove liquid cryogenic agents.
  • C
    Christine Powell MDSep 12, 2025
    What to do if there is audible loss of gas from Bruker Laboratory Equipment?
    • K
      Kyle AndersonSep 12, 2025
      If you hear an audible loss of gas from your Bruker Laboratory Equipment, the membrane or leveling valve of an isolator may be defective, requiring replacement of the leveling valve or isolator. Alternatively, a hose connector may be defective or loose; insert hoses correctly and tighten screws.
  • K
    Kevin DavisSep 18, 2025
    What to do if the ceiling height is too low for assembling Bruker 700'54 Ascend Laboratory Equipment on magnet stand?
    • M
      Mrs. Laura WilliamsSep 18, 2025
      If the ceiling height is too low for assembling your Bruker Laboratory Equipment on the magnet stand because the site does not meet the required conditions, choose another site that meets the required conditions.
  • T
    taylorjessicaSep 5, 2025
    What to do if the magnet system does not reach specification on Bruker 700'54 Ascend?
    • J
      Jane MartinSep 5, 2025
      If the magnet system does not reach specification in your Bruker Laboratory Equipment, it could be due to magnetic material inside the RT bore tube; carefully clean the RT bore tube. Large ferromagnetic parts near the magnet system can also cause this; keep all ferromagnetic parts away from the magnet system as much as possible and repeat shimming.
  • B
    Brian HuntSep 6, 2025
    What to do if the shim current cannot be set correctly on Bruker 700'54 Ascend?
    • K
      kristen02Sep 6, 2025
      If the shim current cannot be set correctly on your Bruker Laboratory Equipment, ensure the control cable is correctly connected to the current lead and to the power supply. Also, check that the switch "Main Coil/OFF/Shim Coil" is in the correct position.
  • N
    Nathan MillerSep 18, 2025
    What to do if the ceiling height is too low for inserting the Helium Level Sensor in Bruker 700'54 Ascend?
    • R
      ryanericSep 22, 2025
      If the ceiling height is too low for inserting the Helium Level Sensor in your Bruker Laboratory Equipment because the site does not meet the required conditions, insert the Helium Level Sensor before mounting the magnet stand.
  • A
    Adam KingAug 30, 2025
    What to do if the helium boil off is too high in Bruker Laboratory Equipment?
    • J
      James WintersAug 30, 2025
      If the helium boil off is too high in your Bruker Laboratory Equipment, it could be because the Helium Level Sensor is in service mode or has been used too often; switch off the Helium Level Sensor and reduce the frequency of helium level measurement. The atmospheric pressure might be decreasing, which is usual behavior, so check the helium boil off daily. Reduced vacuum can also cause this; rebuild the vacuum. Additionally, the radiation baffles might not be inserted in the current lead turret; insert the radiation baffles into the current lead turret.