What to do if RT vessel of Bruker 700'54 Ascend Laboratory Equipment becomes cold and wet?
- CCalvin DawsonAug 31, 2025
If the RT vessel of your Bruker Laboratory Equipment becomes cold and wet, it could be due to a broken vacuum or a vacuum less than 10-3 mbar. Keep the pumping unit connected until filling with liquid helium is finished and continue as in problem Vacuum in RT vessel does not reach 10-6 mbar. Another cause could be a cold leak after transportation. In this case, stop cooling and warm the cryostat.