What to do if the vacuum in RT vessel of Bruker 700'54 Ascend Laboratory Equipment does not reach 5 x 10-5 mbar within 48 hours?
- MMatthew SchroederAug 25, 2025
If the vacuum in the RT vessel of your Bruker Laboratory Equipment does not reach 5 x 10-5 mbar within 48 hours, it could be due to several reasons. The O-rings might be damaged, so check and clean them, replacing if necessary those for the vacuum valve, drop-off plate, reduction and sealing flanges, and the bottom plate. Also, the pumping unit or pumping line could be defective; verify they can reach at least 10-6 mbar with the sealing plug closed, and replace if needed. Scratches or dust on the RT bore sealing surfaces can also cause this, so inspect and clean them. Finally, moisture in the RT vessel can be a culprit; use dry N2 gas to pump and flush the vessel.