What to do if the current lead cannot be removed from Bruker Laboratory Equipment?
- MMr. Harry BoothSep 1, 2025
If you cannot remove the current lead from your Bruker Laboratory Equipment, it is likely that the connector is covered with ice (frozen moisture or nitrogen ice). Carefully remove the ice with warm helium gas over the helium flow system, using the dipstick or the precooling tube as tubing to remove small ice spots from the connector.