ZTKS0164 / Z31949 / 06
Troubleshooting
6.2 Problem
6.2.1 During Transportation
6.2.2 During Assembling
Indicator Possible reason Solution Performed by
Tilt Watch / Shock
Watch activated.
Careless
transportation.
1. Accept delivery with
reservation.
2. Remark the extent of
damage in the trans-
portation documents.
3. Start complaint process.
Approved
Customer
Personnel
Visible damage. Careless
transportation.
1. Accept delivery with
reservation.
2. Remark the extent of
damage in the trans-
portation documents.
3. Start complaint process.
Approved
Customer
Personnel
Indicator Possible reason Solution Performed by
Ceiling height too
low for assembling
on magnet stand.
Site does not meet
the required
conditions.
Choose another site that
meets the required
conditions.
Bruker Service
Ceiling height too
low for inserting the
Helium Level
Sensor.
Site does not meet
the required
conditions.
Insert the Helium Level
Sensor before mounting the
magnet stand.
Bruker Service
Helium bore and
radiation shield are
not centered.
Alignment is not
correct.
Check fitting of the
alignment rods.
Bruker Service
Alignment rod is
loose or broken.
Replace alignment rod
a
.
Bruker Service
Reduction flange is
not centered.
Check the position of the
reduction flange.
Bruker Service