Maintenance
7.3 Maintenance Timetable
7.4 Maintenance-free Components
Interval Device Work Performed by
daily Cryostat Check the helium flow. Approved Customer
Personnel
daily Cryostat Check the nitrogen flow. Approved Customer
Personnel
weekly Cryostat
Check the helium level.
Refill liquid helium if necessary. Follow
the safety instructions of the supplied
manual Refilling Procedure (Z31326).
Record the filling session.
A record of changes in the estimated
consumption may be used for
identification of problems. In this case
contact Bruker Service.
Approved Customer
Personnel
weekly Cryostat
Check the nitrogen level.
Refill liquid nitrogen if necessary. Follow
the safety instructions of the supplied
manual Refilling Procedure (Z31326).
Weekly refill of liquid nitrogen is
recommended.
Record the filling session.
A record of changes in the estimated
consumption may be used to
identify problems. In this case
contact Bruker Service.
Approved Customer
Personnel
Table 7.1: Maintenance timetable
Interval Device Work Performed by
- - - Quench
valves
The quench valves are maintenance-free.
However, in the event of a quench
check the correct position of the globes
in the quench valves.
Approved Customer
Personnel
Table 7.2: Maintenance-free components of the magnet system