General
testing
cond
itions
Radiation
measurements were conducted
Using a Th
ermo micro-rem low energy plastic scintillation radiation survey instrument.
With the survey meter calibrated by the survey meter manufacturer.
Referenced to the center of the detector volume.
Readin
gs
Readin
gs were taken at 50kV, 39 µA, with a Duplex 2205 sample over the examination window
and rou
nded up to the nearest 5 µRem/hr value.
Displayed isodistance contours show the maximum value likely to be encountered in x10
-6
Rem/
hr
(rem/hr) or equivalently in x10
-8
Sieverts
/hr.
Duplex
μrad(μSieverts)/hr
Location
10cm 30cm 100cm
A ≤45(0.45)
≤20(0.20)
≤5(0.05)
B
≤10(0.10)
Background
Background
C
≤5(0.05)
≤5(0.05)
Backgroun
d
D
Backgroun
d
Backgroun
d
Backgroun
d
E
Backgroun
d
Backgroun
d
Backgroun
d
F
Background
Background
Background
G
Backgroun
d
Backgroun
d
Backgroun
d
H ≤40(0.40)
≤5(0.05)
Backgroun
d
I ≤40(0.40)
≤5(0.05)
Backgroun
d
J
≤10(0.10)
Backgroun
d
Backgroun
d
K
≤10(0.10)
Backgroun
d
Backgroun
d
L
≤5(0.05)
Background
Background
M
≤5(0.05)
Backgroun
d
Backgroun
d
Your Complete Source for
Testing Equipment
www.BergEng.com
Berg Engineering & Sales Company, Inc.
1-847-577-3980
Info@BergEng.com

Other manuals for Bruker S1 TITAN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker S1 TITAN and is the answer not in the manual?

Bruker S1 TITAN Specifications

General IconGeneral
BrandBruker
ModelS1 TITAN
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals