testing 
cond
itions 
Radiation 
measurements were conducted 
– 
 Using a Th
ermo micro-rem low energy plastic scintillation radiation survey instrument.
 With the survey meter calibrated by the survey meter manufacturer.
 Referenced to the center of the detector volume.
Readin
gs were taken at 50kV, 39 µA, with a Duplex 2205 sample over the examination window 
and rou
nded up to the nearest 5 µRem/hr value. 
Displayed isodistance contours show the maximum value likely to be encountered in x10
-6
Rem/
hr 
(rem/hr) or equivalently in x10
-8 
Sieverts
/hr.
Duplex 
μrad(μSieverts)/hr 
Location
10cm  30cm  100cm 
A  ≤45(0.45)
≤20(0.20)
≤5(0.05)
C 
≤5(0.05)
≤5(0.05)
Backgroun
d
H  ≤40(0.40)
≤5(0.05)
Backgroun
d
I  ≤40(0.40)
≤5(0.05)
Backgroun
d
L 
≤5(0.05)
Background
Background
M 
≤5(0.05)
Backgroun
d
Backgroun
d