12. Quali
ty Check
This tes
t verifies assays measured and recorded 
on the instru
ment are accurate.
 
To verify
 that your instrument is
 calibr
ated 
corr
ectly
 –  
Set APPLI
CATION
 to the o
ption applicable to the check sample supplied with the 
instru
ment and Method to Automatic Calibration Selection. 
From th
e 
Read
y to Test
 screen, 
measure the supplied 
check sample 1
0 times.
 
Average the 10 assays. See Averaging, page 58. 
Compar
e the average measured value 
for each 
element against the Acceptance Limit 
maximum
 and minimum on the Check Sample Report supplied with your instrument. 
If all av
erage
d measurem
ents fall within MAX and MIN limits, continue to use the 
instru
ment. 
If averages fall outside MAX and MIN limits, with respect to standard deviation, check – 
 Measurement times
 Sample surface
 Instrument window
If the instrument does not perform as expected, contact Bruker at 
support.hmp@bruker.com, +1 (509) 783-9850. 
reports
 
The followi
ng figure is an example only.
 When p
erforming a quality check, use the 
Check
 Sample 
Report inclu
ded with your instrument. It contains values specific to it as determined during 
testing at the factory. 
Check Sample Report, example only 