S1 TI
TAN, T
RACER
5, and CTX
User Manual
030.0112.04.0
29 of 67
factors
During tes
ting, the following correction factors were
used:
Distance
Nist H30 Correction Factor
5 cm
3.612
10 cm
3.612
30 cm
2.948
100 cm
2.919
dose in mRad
In mRad
Worst-Case 1 Highest
Energy
Standard Setting 1
Alloy Phase 1
Standard Setting 2 Alloy
Phase 2
50Ke
V, 39A 40KeV, 5.2A 15KeV, 11.35A
Blank, No
Filter Ti25,Al300m filter Blank, No Filter
mRad mRad mRad
Distan
ce
(cm)
Open
Beam With Sample
Open
Beam With Sample
Open
Beam With Sample
100 1839
Background 120 Background 34 Background
30 3320
0.027 1126 Background 398 Background
10 143697 0.220 9031 Background 2854 Background
5 163420
0.372 10656 Background 3161 Background
Surface 218936 0.632 14006 4310
Eye / Skin
T
ime to
Limit
4.2 / 13.7
Minutes
2.7 / 9
Years
66 / 214
Minutes
214 / 696
Minutes
dose in
mSieverts
In
mSiev
erts
Worst-case 1 Highest
Energy
Standard Setting 1
Alloy Phase 1
Standard Setting 2 Alloy
Phase 2
50Ke
V, 39A 40KeV, 5.2A 15KeV, 11.35A
Blank, No
Filter
Ti25,Al300m filter
Blank, No Filter
mSi
everts mSieverts mSieverts
Distance
(cm)
Open
Beam With Sample
Open
Beam With Sample
Open
Beam With Sample
100 18.39
Background 1.197 Background 0.339 Background
30 33.20
0.000265 11.262 Background 3.98 Background
10 1437.0
0.002204 90.308 Background 28.537 Background
5 1634.2
0.003721 106.563 Background 31.608 Background
Surface
2189.4 0.006324 140.06 43.100
Eye / Skin
Time to
Limit
4.2 / 13.7
Minutes
2.7 / 9
Years
66 / 214
Minutes
214 / 696
Minutes