5 | Installation BÜCHI Labortechnik AG
20/42 Operation Manual Recirculating Chiller F-305 / F-308 / F-314
5 Installation
5.1 Important considerations prior to commissioning
DANGER
Risk of fatal injury if used in potentially explosive atmospheres
u Do not store or operate the device in potentially explosive atmospheres.
u Do not bring the device into close proximity with flammable vapors.
u Do not place any open solvent containers near to the device.
WARNING
Risk of fire from overheating device
u Do not cover over the device.
u Make sure there is adequate air circulation.
u Position the device at least 40 cm away from other objects.
CAUTION
Risk of injury if device falls or is dropped
u The device should be lifted carefully and by two persons.
u Do not drop the device or shipping crate.
u Place the device on a firm, level and vibration-free surface.
5.2 Installation site
The installation site must meet the following requirements:
Firm and level surface
Min. space requirement 500 mm x 600 mm (W x D)
Clearance on all sides: at least 40 cm
Adequate air circulation
Power supply outlet easily accessible (even after device has been set up)
The Recirculating Chiller is intended for use in laboratory environments (see Chap-
ter3.3.2 "Ambient conditions", page17).