BÜCHI Labortechnik AG Operation | 6
Operation Manual Recirculating Chiller F-305 / F-308 / F-314 31/42
6.4 Operating chiller with Interface I-300/I-300 Pro
The Recirculating Chiller can also be externally controlled by an
Interface I-300/I-300 Pro.
u Switch on the device by means of the On/Off switch (1).
u Connect the Interface I-300/I-300 Pro (see Chapter5.4.2 "Connecting the
communication cable to the F-3xx", page24) and the relevant sections of the
instruction manual for the I-300/I-300Pro.
u Set the required temperature via the Interface I-300/I-300Pro, see operating
instructions for I-300/I-300Pro.
u Start/stop the chilling process via the Interface I-300/I-300Pro, see operating
instructions for I-300/I-300Pro.