4 Description of function
31
NIRFlex N-500 Operation Manual, Version G
4.6.4 Vial add-on for NIRFlex Solids
NIRFlex Solids with vial add-on
The NIRFlex Solids with the vial add-on is the ideal configuration to measure solid samples in glass
vials. This special add-on is equipped with 6 positions for sample vials with a diameter of
10−15 mm. Buchi provides one type of glass vials as an optional accessory for this specific add-on.
This built-in External Reference is measured automatically without any interaction of the operator.
A measurement sequence consists of a measurement of the reference and afterwards the mea-
surement of the sample positions (1−6), which have been pre-defined in the NIRWare Operator. The
actually measured spectrum is set against the last measured reference.
4.6.5 Tablet add-on for NIRFlex Solids
NIRFlex Solids with tablet add-on
The NIRFlex Solids with the tablet add-on is the ideal configuration to measure tablets in diffuse reflec-
tion mode. This special add-on is equipped with 10 sample positions for round tablets with a diameter
between 5−10 mm.
This built-in External Reference is measured automatically without any interaction of the operator.
A measurement sequence consists of a measurement of the reference and afterwards the measure-
ment of the sample positions (1−10), which have been pre-defined in the NIRWare Operator.