BÜCHI Labortechnik AG Operation | 6
Operation Manual Rotavapor® R-100 29/60
u Switch off the Rotavapor on the heating bath.
u Hold the condenser (1) firmly with one hand and release the lever (2) with the
other hand.
u Adjust the immersion angle.
u Lock the rotary drive unit in position again with the lever (2).
u Make sure that there is at least 10 mm clearance between the edge of the heating
bath and the evaporating flask or vapor duct. To achieve that, adjust the height of
the rotary drive unit if necessary (see Chapter6.2.1 "Raising and lowering the
evaporating flask", page33).
u Check the vertical end stop and adjust if necessary (see Chapter6.1.5 "Vertical
adjustment", page30).