VibroMet
®
2
Vibratory Polisher

Questions and Answers

  • H
    Heather SanchezAug 4, 2025
    Why are the specimen holders not rotating at the normal rate on my Buehler Vibromet 2?
    • T
      Travis WuAug 4, 2025
      If your Buehler Laboratory Equipment specimen holders are not rotating at the normal rate of 6 revolutions per minute, first, try increasing the rate if the control is set too low. Also, ensure that you add clean polishing medium if there is insufficient lubricant. Check that the cloth is secured properly, and level the machine to ensure correct operation.
  • A
    Alvin HamiltonAug 10, 2025
    What to do if my Buehler Laboratory Equipment is making a loud striking noise?
    • J
      johnmerrittAug 11, 2025
      If your Buehler Laboratory Equipment is making a loud striking noise, it could be due to the air gap not being adjusted correctly. Adjust the air gap to the specified .035 inch. Another cause could be the absence of lubrication between the weights and holders. To resolve this, coat the contacting surfaces with a clean lubricant.
  • G
    gabrielle42Aug 16, 2025
    How to troubleshoot a Buehler Laboratory Equipment with no response?
    • J
      jasminecummingsAug 16, 2025
      If your Buehler Laboratory Equipment is not responding, the issue might stem from a faulty switch. Consider replacing or repairing the switch. Alternatively, a failed magnet could be the culprit, in which case, replace the magnet. Finally, an electrical short could be the cause, and you should repair the short.
  • A
    Ashley RiversAug 22, 2025
    What causes weak action in Buehler Vibromet 2?
    • V
      vazquezkatherineAug 22, 2025
      If your Buehler Laboratory Equipment has weak action, it could be due to low voltage. Correct the power supply to address this. Alternatively, a loose bowl could be the cause; tighten the securing screws. Broken springs may also lead to weak action, so replace the springs if needed. Finally, loose springs could be the issue; tighten the spring bolts.
  • T
    Timothy BrownAug 27, 2025
    What to do if Buehler Vibromet 2 Laboratory Equipment hums but has no vibration?
    • J
      Jamie ColeAug 27, 2025
      If your Buehler Laboratory Equipment is humming but not vibrating, the issue could be a defective SCR or rectifier. Replace the broken parts to resolve this problem.

Summary

EC - DECLARATION OF CONFORMITY

Compliance Standards

Lists the Safety of Machinery and EMC standards the product adheres to.

Warranty Information

Warranty Coverage and Terms

Outlines the warranty period, conditions, and coverage for defects.

Return for Warranty Service

Instructions and required information for returning the unit for warranty service.

Product Specifications

Dimensional and Electrical Specifications

Provides physical dimensions, voltage range, and wattage of the polisher.

Unpacking and Initial Handling

Safety Information and Precautions

General Safety Precautions

Essential safety guidelines for operators, including protective gear and environment.

Installation Procedures

Electrical Installation Safety

Crucial safety warnings and procedures for electrical installation.

Operation and Procedures

Recommended Polishing Procedures

Guidelines for polishing various specimen materials with specific abrasives and times.

Maintenance and Troubleshooting

Adjusting the Air Gap

Detailed procedure for setting the correct air gap for optimal performance and preventing damage.

Troubleshooting Common Issues

A chart listing problems, possible causes, and solutions for the VibroMet 2.