32
8045
ENGLISH
TEST MENU
ENTER
ENTER
ENTER
0......9
12 s
0......9
ENTER
ENTER
4.5.4 FLOW SIMULATION
A flow value can be simulated within this menu, allowing the user to test the system without
any liquid being present. The simulated value influences all the outputs including the
relays and pulse output.
Enter flow value
Press
0......9
or to stop the flow simulation.
4.5.3 CALIBRATION OF THE ZERO POINT
Fill the pipe with the measured fluid and stop the flow. To calibrate the unit, press
enter when CALIB 0 is diplayed within the test menu and select CALIB Y. After selection
the transmitter will automatically set the zero-point after 12 seconds.
The sensor must be immersed in fluid 24hrs before calibration. Ensure
there are no air bubbles in the pipe and the fluid is not moving before
commencing the calibration.
This calibration is only valid for the actual parameters (pipe, fitting and fluid
characteristics) and must be completed before the determination of the
K-Factor via the Teach-In method.
CALIB 0
CALIB N
CALIB Y
MEASURE
FLOW
FLOW
4.5.3