GasAlertMicroClip
Alarms
9
Alarm Display Alarm Display
Sensor Alarm:
Displays Err
Automatic Shutdown Alarm:
Eight beeps and eight flashes
LOW BAT and L display
Vibrator alarm temporarily activates
Displays OFF before turning off
Low Battery Alarm:
(Confidence beep disabled)
One beep and one flash every 5
seconds
and L flashes
Confidence Beep:
One beep every 10 seconds
Note
Alarms can be set to be latching or non-latching. To confirm this setting, access the latching alarms option in the GasAlertMicroClip
Soft Tools software. Local regulations may require a latching alarm.
If the detector is in stealth mode, it only vibrates in alarm mode (the audible and visual alarms are disabled). If the detector is in IR
stealth mode, the IR LEDs and vibrator activate in alarm mode (the audible alarm is disabled).
If the Low Alarm Acknowledge option is enabled, the beeper can be disabled only during a low alarm by pressing C. The vibrator
and LEDs will function normally. If the alarm escalates to a high, STEL, or TWA alarm, the beeper reactivates.
The backlight activates during any alarm condition.