8555-M512-O1 Rev D 01/15 25
ZERO
With ZERO displayed, press the PRINT/ENTER key. The display will change to ZERO LIMIT.
Proceed to the Two Percent Zero Tracking Range Limit parameter.
Otherwise, press the ID/HEIGHT key to advance to the FILTER menu.
ZERO LIMIT (Two Percent Zero Tracking Range Limit)
This setting limits the scale to zero only + or - 2% of the scale capacity.
If enabled, (ZERO LIMIT = YES) zero tracking and push button zero is limited to 2% of
scale capacity.
If disabled, (ZERO LIMIT = no) zero tracking and push button zero will operate up to full
scale capacity.
The display will show ZERO LIMIT and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO
or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it.
Allowable values are YES or no.
YES = 2% Zero Limit is Enabled no = 2% Zero Limit is Disabled
POWER UP ZERO (Power Up Zero)
The display will show POWER UP ZERO and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO
or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it.
Allowable values are YES or no.
YES = Weight display will be reset to
zero automatically on power up
no = Weight display will not be reset to
zero on power up
ZERO TRACKING (Zero Tracking)
The zero tracking is a value in scale divisions that will automatically zero the scale within the
limits determined by the setting for this prompt.
The display will show ZERO TRACkING and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 0 through 18 (1 to 9 divisions by 0.5 divisions) e.g. 1 = .5 grads, 2 = 1
grad, 3 = 1.5 grads, etc…
Note that entering a 0 (zero) will disable zero tracking.