72 Infant Flow
®
SiPAP
675-120 Revision C November 2004
Removal and Fitting of Case Bleed Filtered
Restrictor
Refer to Figure 18.
Removal
1 Remove the case.
2 Remove the case bleed filtered restrictor (3) from the pipe.
3 Discard the case bleed filtered restrictor (3).
Fitting
1 Install the new case bleed filtered restrictor (3).
2 Screw on the water trap bowl (1) and tighten using the Special Spanner.
3 Install the case.
Figure 18 Water Trap Filter Removal/Fitting