36
Calibration Check
Connect a 3 litre syringe to the
transducer with the adapter supplied and
empty by pushing the handle fully in.
Select ‘Calibration Check’ from the main
menu and then select ‘check Calibration’.
Fill the syringe by pulling the handle at a
constant rate until the end stop is
reached and then immediately empty the
syringe completely.
Press ‘Done’ to view the calibration check
results.
Press ‘Again’ to repeat the calibration
check at a low flow rate.
Press ‘Again’ to repeat the calibration
check at a high flow rate.
When a calibration check at all three flow
rates has been completed press ‘Done’ to
view the calibration check report screen.