37
The calibration error for expiration and
inspiration are displayed. The calibration
error should be less that 3%. If a
greater error is shown, repeat the
procedure ensuring that the syringe is
emptied and filled in a smooth manner
without jerking the handle. If an error
greater than 3% is still shown, inspect
the turbine transducer and clean if
necessary.