Storage Storage Temperature: 10 - 40˚ C (50˚ F to 104˚ F)
Storage Relative Humidity 30% - 75%, non-condensing
Additional Information Important Note: THE CABLE AND PROBE SHOULD BE COVERED BY
A STERILE DRAPE WHEN USED IN SURGERY.
The probe cable may be used in a sterile field after a sterile drape is placed over the probe and
cable. Sterile drapes are available from medical supply companies that will allow use with a trocar
sheath or cannula.
The suggested installation sequence of a sterilised probe into a sterile drape is as follows:
a) A professional in the sterile field (i.e. a scrub nurse) holds the sterilized
OmniProbe
®
device and the sterile drape.
b) A professional outside the sterile field (i.e. a circulating nurse) holds the cable.
c) The scrub nurse inserts the probe into the drape and passes the rear end of the drape
(without pulling it out to its full length) to the circulating nurse.
d) The circulating nurse takes the rear end of the drape in one hand and the cable
in the other hand, reaches into the un-extended drape with the cable and inserts
the cable connector into the OmniProbe connector.
e) The circulating nurse then pulls the drape up the cable to its full length,
and connects the cable to the C-Trak Analyser.
Manufacturer Contact Southern Scientific Ltd. (Europe and RoW)
E-mail: info@southernscientific.co.uk
Telephone: +44 1273 497600
LabLogic Systems, Inc. (USA)
E-mail: sales@carewise.com
Telephone: +1 813 626 6848
The instructions provided above have been validated by the manufacturer of the medical
device as being capable of preparing a medical device for reuse. It remains the responsibility
of the processor to ensure that the processing, as actually performed using equipment,
materials and personnel in the processing facility, achieves the desired result. This requires
verification and/or validation and routine monitoring of the process.