10.2.3 Calibration Check With Built-in Weight
(CAUW-D/CAUW/CAUX series only)
“i-tESt” (The balance is checked with the built-in weight but not adjusted.)
1 From the mass display, press the [CAL] key
once. If the display shows “i tESt”
(Preset calibration method is “i-tESt”),
jump to Step 5.
If the display shows another item
(Preset calibration method is not “i-tESt”),
go on to Step 2.
(Refer to 10.3.1 for preset calibration method)
2 Press the [CAL] key repeatedly until
“FUnC.SEL” appears, then, press the [O/T]
key to display“CAL”.
3 Press the [O/T] key again. The display shows
“E CAL”.
4 Press the [CAL] key repeatedly until the
display shows “i tESt”.
5 Press the [O/T] key. The display changes
automatically in this order: “tESt 2”, “tESt 1”,
and “tESt 0”. After that, the display shows the
drift from the previous span calibration for
several seconds.
6 After “tEStEnd” appears for several seconds,
the display returns to the mass display, and the
calibration test is complete.
Note that the Calibration Test procedure DOES
NOT adjust the balance, it only reports the drift
from the previous span calibration.
...