11
6
Cornerload Performance
1. Allow the balance to warm up sufficiently by
turning ON the power and leaving it at the
gramdisplay at least two hours before starting
the performance checks.
2. Use a weight that is approximately 1/4 of the
balance capacity and place it sequentially at five
different positions on the pan in the order
shown. Record the readings in each of the five
positions.
3. If the difference between the readings at the
center position and each of the off-center
positions (eccentric error) is within five counts,
the balance cornerload performance is
considered normal.
* One count corresponds to one minimum display
unit of the balance model.
A-7. Below-Weigh Hook Dimensions
(unit : mm)
0.7
4.5
Below-weigh Hook
Hole 20
Balance bottom
43
3
4.6