CAS CUW Series - Calibration Setting; Selecting the Calibration Type; PSC Fully-Automatic Calibration (UW Series Only)

CAS CUW Series
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52
10.3 Calibration Setting
10.3.1 Selecting the Calibration Type
Set the calibration type that will be used in Calibration Execution.
To set up “Span calibration using the built-in weight”,(UW only) Select menu item 1 .
To set up “Calibration check using the built-in weight”,(UW only) Select menu item 2 .
To set up “Span calibration using external weights”, Select menu item 3 .
To set up “Calibration check using external weights”, Select menu item 4 .
10.3.2 PSC Fully-automatic Calibration (UW series only)
With the PSC function, span calibration is performed automatically using the built-in calibration
weight when the balance detects a temperature change that would affect weighing accuracy.
1. To turn ON the PSC function, Select menu item
5 .
2. To turn OFF the PSC function, Select menu
item 6 .)
x Blinking calibration symbol indicates an automatic calibration is about to start.
x If PSC starts while the balance is in use, press the [POWER] key to abort that cycle.